Development of a burst wave lithotripsy system for noninvasive fragmentation of ureteroliths in pet cats

Abstract

Background

Upper urinary tract stones are increasingly prevalent in pet cats and are difficult to manage. Surgical procedures to address obstructing ureteroliths have short- and long-term complications, and medical therapies (e.g., fluid diuresis and smooth muscle relaxants) are infrequently effective. Burst wave lithotripsy is a non-invasive, ultrasound-guided, handheld focused ultrasound technology to disintegrate urinary stones, which is now undergoing human clinical trials in awake unanesthetized subjects.

Results

In this study, we designed and performed in vitro testing of a modified burst wave lithotripsy system to noninvasively fragment stones in cats. The design accounted for differences in anatomic scale, acoustic window, skin-to-stone depth, and stone size. Prototypes were fabricated and tested in a benchtop model using 35 natural calcium oxalate monohydrate stones from cats. In an initial experiment, burst wave lithotripsy was performed using peak ultrasound pressures of 7.3 (n = 10), 8.0 (n = 5), or 8.9 MPa (n = 10) for up to 30 min. Fourteen of 25 stones fragmented to < 1 mm within the 30 min. In a second experiment, burst wave lithotripsy was performed using a second transducer and peak ultrasound pressure of 8.0 MPa (n = 10) for up to 50 min. In the second experiment, 9 of 10 stones fragmented to < 1 mm within the 50 min. Across both experiments, an average of 73–97% of stone mass could be reduced to fragments < 1 mm. A third experiment found negligible injury with in vivo exposure of kidneys and ureters in a porcine animal model.

Conclusions

These data support further evaluation of burst wave lithotripsy as a noninvasive intervention for obstructing ureteroliths in cats.

Peer Review reports

Background

Upper urinary tract stones in cats are a significant cause of morbidity and mortality [1, 2] The vast majority of these stones (87–98%) are calcium-based [3,4,5]. There is no protocol for medical stone dissolution of calcium-based stones, and medical management with intravenous fluid therapy and drugs to facilitate obstructing stone passage are effective in only a small proportion of patients (13%) [1, 6]. In the remaining cases, placement of a subcutaneous ureteral bypass (SUB) device or ureteral stent is recommended to relieve obstruction [7]. However, these procedures carry a 6–18% perioperative mortality rate and a 29–56% risk of long-term complications such as device occlusion, urinary tract infection, and, in the case of stents, lower urinary tract signs without infection [4, 8,9,10,11]. The high costs and specialized techniques required for placement of SUB devices or stents also limit availability of these options. For these reasons, new minimally invasive techniques and procedures are sought to improve the efficacy, safety, and availability of treatments for feline ureteroliths.

Minimally invasive techniques such as shock wave lithotripsy (SWL) [12] and endoscopic laser lithotripsy [13] are the primary interventions for stones in humans. These methods are not available for treating ureteroliths in cats, as feline calcium oxalate uroliths are relatively resistant to fragmentation by SWL and the feline ureter is too small for laser lithotripsy instrumentation [7, 14]. Burst wave lithotripsy (BWL) is a new noninvasive approach to fragment urinary tract stones based on focused ultrasound technology. This method uses a focused ultrasound transducer to apply short harmonic bursts of ultrasound to a stone to produce cyclic stressing that leads to fractures and fragmentation [15]. A BWL system consists of a small electronic pulser and therapy transducer that can be coupled to the skin of a patient by a thin layer of ultrasound gel. An ultrasound imaging probe incorporated with the therapy head allows detection and localization of a stone, using custom algorithms to enhance detection of calcifications [16, 17]. This technology has progressed from conception to human clinical trials over the last 8 years [18, 19]. BWL has been demonstrated to fragment a number of different stone compositions, including calcium oxalate stones, in preclinical and human studies [19, 20], without producing any significant injury or complications [20, 21]. In a preclinical study using calcium oxalate monohydrate stones (known to be resistant to SWL) implanted in the kidneys of pigs, stones were noninvasively fragmented by BWL, with 88% of the resulting fragments smaller than 2 mm [20]. No injury was found to the kidney parenchyma and only mild hemorrhagic injury to the collecting space lining. The fine control of acoustic parameters such as amplitude and frequency has been demonstrated to have a number of benefits, one of which is the control of the size of fragments generated from the stone during the procedure [19]. For instance, 170 kHz ultrasound bursts produce fragments in artificial stones up to 4 mm, while 800 kHz ultrasound produce fragments approximately 0.6 mm in size. While the normal luminal diameter of the feline ureter is estimated at 0.3–0.4 mm, the average diameter of an obstructed ureter is 3.2–3.5 mm, and it can dilate up to 11 mm [2, 8, 22]. Furthermore, the typical diameter of ureteral stents in cats is 1 mm, and only 2% of ureteroliths are ≤ 1 mm in cats [4, 11, 22]. BWL can therefore be designed to noninvasively fragment feline ureteroliths to pieces small enough to pass through the cat ureter, while minimizing the risks of injury and complications associated with surgery.

In this paper, we describe the design and testing of a BWL system adapted to noninvasively treat ureteroliths in cats. These studies include analysis of the acoustic window and skin-to-stone depth in cats, design of electronics and a small ultrasound-guided transducer, demonstration of stone fracture on natural feline ureteroliths, and preliminary evaluation of potential soft tissue injury.

Methods

Burst wave lithotripsy uses a focused ultrasound transducer to apply high-amplitude pressure pulses to a stone to fragment it. However, human systems that have been designed for this application are not ideal for treating cats, as the therapy devices are too large, must focus deep (5-10 cm) from the skin surface, incorporate imagers for abdominal imaging in humans, and are designed to fragment stones into ≤ 2 mm fragments (small enough for humans to pass but not necessarily for cats). To address these differences, we have developed an initial prototype system designed for use in cats, operating at 650 kHz with a smaller transducer to treat at shallower depths than humans and adapted for a veterinary ultrasound imager. Computed tomography and ultrasound images of cats with ureteroliths were analyzed to determine the optimal acoustic window and the appropriate dimension for the transducer. The transducer is designed to target ureteroliths while avoiding bone and other vital structures (Fig. 1).

Fig. 1
figure 1

A Three-dimensional anatomic surface models derived from CT data, showing the urinary tract, bone structure, and intestinal tract in a cat. B Surface rendering of the skin, with green points indicating center positions of a 5-cm diameter transducer with a clear acoustic window to the left kidney, left proximal ureter, and urethra (top to bottom). The pink-yellow scale indicates the skin-to-stone depth at different points along the surface with a clear acoustic path to the stone

Anatomic analysis

Image analysis from B-Mode ultrasound images and computed tomography (CT) 3-dimensional data were performed to assess an approximate focal depth and acoustic window for design of the BWL transducer. The acoustic window is the anatomic aperture where ultrasound can pass through soft tissue to the focal point without encountering reflective obstructions such as bone or gas in the intestines or lung. This window limits the dimensions of the ultrasound device placed on the skin.

Images were captured from previous patients with ureteral stones. Skin-to-stone distance was measured in ten cats manually from ultrasound images. CT data for one cat was manually outlined using custom software [23] to segment the skin surface, bone, kidneys, ureters, bladder and urethra, and small/large intestine as separate tissues. The outlines were converted to 3D surface models using custom software developed in MATLAB (MathWorks, Natick, MA). An acoustic window analysis was performed to determine the largest-diameter ultrasound transducer that could be applied without bone or intestines (which block ultrasound) intervening in any part of the beam path to a target location within the urinary tract. Target locations in the kidney, proximal and distal ureter, and urethra were assessed. The distances at these different locations from the transducer were also calculated as a function of the transducer position to determine optimal windows for therapy in each scenario.

Device design and fabrication

Based on image analysis, a BWL transducer was designed using a circular piezoelectric transducer element and lens model [24]. The transducer beam profile was calculated using the Rayleigh integral [25, 26] and custom piezoelectric simulation software based on a version of the KLM model [27]. The frequency of the ultrasound was selected based on relative frequency to reduce stones to a size < 1 mm [19, 20].

The transducer was designed in a solid modeling program (Solidworks, Dassault Systemes, Vélizy-Villacoublay, France). The housing incorporating the focusing lens was produced on a resin-based 3D printer (3SP, Envisiontech Dearborn, MI) and the piezoelectric elements were bonded to the lens through a quarter-wavelength epoxy-aluminum oxide composite acoustic matching layer. A water bolus with latex rubber membrane was incorporated to the front of the transducer to couple the transducer to skin. Two fluid ports were installed to allow degassed water to be added to the space between the membrane and transducer for coupling. An RG-174 coaxial cable was soldered to the transducer element to apply voltage to produce ultrasonic vibration of the piezoelectric element.

A small electronic pulser was constructed to apply the voltage signal to the transducer and power it. The pulser was constructed based on a previous design to apply short, high amplitude sinusoidal voltage pulses [28]. The system used an FPGA based architecture to be able to rapidly reconfigure the output parameters as needed during testing. A passive electrical network was constructed to increase the voltage applied to the transducer and the power output.

The system was calibrated using a fiber optic probe hydrophone (FOPH2000, Leutenbach, Germany) in an open bath of degassed, deionized water [29]. The transducer was positioned facing the hydrophone probe tip and the peak pulse pressure values, focus location, and focus beamwidths were measured by scanning the probe tip while affixed to a 3-axis motorized positioner. The collected signals were output to an oscilloscope (DSOX3034, Keysight, Santa Rosa, CA) and deconvolved using custom software programmed in MATLAB.

Testing on stones

To assess the capability of the transducer to fragment uroliths, the assembled BWL system was tested on natural stones. Calcium oxalate stones (100% monohydrate) were collected from open surgeries at University of Minnesota Veterinary Medical Center. The stone diameters were between 2–5 mm. Stone composition was analyzed by polarization microscopy and infrared spectroscopy and shipped in a dry state. The stones were rehydrated in water at least 48 h before experiments.

A small plastic holder was used to position each stone at the focus of the transducer in a degassed water bath. Each stone was held freely in a small depression in a molded synthetic gelatin mass at the center of the holder to mimic the surrounding tissue. The holder magnetically attached to the front of the transducer to automatically position the stone at the center of the focus.

Experiments were performed in two stages. A first experiment was conducted to compare three different pressure amplitudes for exposure and to optimize this parameter. Burst wave lithotripsy pulses were applied at focal pressure amplitudes of 7.3, 8.0, or 8.9 MPa with 20-cycle pulse duration at a rate of 10 pulses per second. Each stone was treated in 10-min intervals for up to 30 min total exposure. At each interval, the stone fragments were passed through serial sieves with 1.0-mm and 0.7-mm openings. The fragments larger than each size were weighted with an analytical balance and replaced in the holder for further treatment to determine the fraction of stone fragmented at each time point.

A second experiment was conducted with the pressure level showing the greatest efficacy (8.0 MPa) to confirm the consistency of the results with a second transducer. The BWL parameters were kept the same, but stones were treated for up to 50 min in 10-min intervals, which we considered clinically practical.

In vivo evaluation in a porcine kidney and ureter

All procedures were approved by the University of Washington’s Institutional Animal Care and Use Committee. In the third experiment, 3-mm calcium oxalate monohydrate stones were soaked overnight before being implanted into the kidneys and ureters of four live pigs as described previously [30]. The pig was chosen due to our experience working with the pig model for stone comminution and expertise in evaluating potential injury from BWL in this model. Pigs were obtained from Progressive Swine Farms (Woodinville, WA), an approved swine provider to the University of Washington. Ties were placed on either side of the stone in the ureter to prevent movement of the stone during treatment. The procedure was performed in an open abdomen, with the transducer directly coupled to the kidney/ureter by degassed saline and placed over the stone. As the ultrasound imaging was unavailable at the site of animal studies, an audible clicking associated with the BWL bursts impacting the stone confirmed alignment and was used to realign the focus if the stone or transducer moved. Each stone was treated for 10 min at 8 MPa. After treatment, the location of the transducer was marked to ensure tissue was harvested from the correct location. After treatment, the pig was euthanized with an intravenous injection of sodium pentobarbitol (87 mg/kg, Euthasol ®, Virbac Inc., Westlake, TX) while still under anesthesia (1–3% isoflurane), according to AVMA guidelines. After euthanasia, the kidneys and ureter were harvested for gross evaluation before being processed for histological evaluation. Tissue sections taken from the treatment locations were stained with hematoxylin and eosin (H&E).

Results

Acoustic window analysis

Ten B-Mode ultrasound scans were analyzed to determine a range of skin-to-stone depths for obstructing ureteral stones. All stones were easily visualized by ultrasonography. The scans indicated depths ranging from 0.7 – 2.1 cm from the skin surface, with a mean of 1.3 cm.

CT analysis showed minimum skin-to-stone distances from 2.4 to 3.8 cm for seven locations (Table 1), somewhat greater than the range captured by ultrasonography. However, the distances calculated from CT do not take into account the compression of tissue typically applied by the probe to maintain contact during ultrasound imaging. The analysis indicated that all stone locations could be targeted with an open acoustic window for a transducer maximum size up to 6 cm. The available transducer positions and skin-to-stone depths for 3 stone locations for a 5-cm diameter transducer are shown in Fig. 1. The minimum acoustic window size is defined as the spatial distribution of allowed transducer center-points (green points in Fig. 1B).

Table 1 Skin-to-target depths and acoustic window minimum dimensions for a simulated transducer calculated from computed tomography images of a cat as a function of location

Transducer design

The transducer was designed with a therapy aperture of 3.3 cm diameter, with a focal length of 3.7 cm. The transducer was constructed from a flat piezoceramic element bonded to a focusing lens through an acoustic matching layer. The distance of 3.7 cm focal length is from the apex of the concave lens to the center of the focus, which placed the front edge of the transducer approximately 1.6 cm from the focus when against the skin surface. To accommodate shallower skin-to-stone distances, a flexible membrane was fitted over the transducer surface. The gap between the membrane and transducer is filled with water to cause the membrane to expand and adjust the position of the focus relative to skin surface. Coupling gel is applied first to ensure good ultrasound transmission at the membrane-skin interface. A microconvex imaging transducer with bandwidth of 4–10 MHz (PVT712, Toshiba Medical, Otawara, Tochigi, Japan) was incorporated, obliquely aligned the therapy transducer (Fig. 2). With the combined therapy and imaging transducers in position, the overall aperture of the device was 5.0 cm (Fig. 3).

Fig. 2
figure 2

A Simulated ultrasound pressure field for the designed burst wave lithotripsy transducer. The pressure value corresponds to the relative pressure between the focus and surface of the transducer. B Cross sectional diagram of the transducer design, containing the piezoelectric element, acoustic lens, and off-axis ultrasound imaging transducer for therapy monitoring

Fig. 3
figure 3

Design of a burst wave lithotripsy system for fragmentation of feline ureteroliths. A The fabricated transducer and pulse generator with power supply. B Close-up of the transducer applicating surface, using a rubber membrane covering filled with water between the membrane and transducer for flexible contact with the skin

The transducer beam geometry was characterized under linear (low amplitude) acoustic output. The focal region was ellipsoidal with a lateral -6 dB beamwidth of 3 mm and an axial beamwidth of 17 mm. Hydrophone measurements recorded a peak negative pressure output of up to -8.9 MPa at the maximum driving voltage. Comparatively, human treatments have been performed at up to 7 MPa in situ pressure amplitude, suggesting sufficient pressure to fragment stones [18, 31, 32].

Stone exposure

All stones were composed of 100% calcium oxalate monohydrate and had nearly spherical morphology. The stones were between 2–5 mm maximum dimension and had a mass between 4 – 61 mg (average mass 15–16 mg for each group). In the first experiment, the stones were exposed to either 7.3 MPa (n = 10), 8.0 MPa (n = 5), or 8.9 MPa (n = 10) peak ultrasound pressure in 10-min increments up to 30 min total exposure per stone. Figure 4 shows the average fragmented mass as a function of time for each of the 3 pressure levels. Depending on the pressure level administered, the average stone mass reduced to pieces smaller than 1 mm varied from 73 ± 32% to 97 ± 5%, and between 43 ± 28%—77 ± 19% reduced to fragments smaller than 0.7 mm after 30 min (Fig. 5). Of the 25 stones treated, 14 (56%) were completely reduced to fragments < 1 mm within 30 min of exposure. 2–3 stones in each group achieved complete fragmentation to < 1 mm pieces within 10 min. Only 2 of the treated stones showed no visible fragmentation over the 30-min period.

Fig. 4
figure 4

Example of a feline calcium oxalate urolith before (left) and after (right) 10 min of burst wave lithotripsy at 8.0 MPa pressure amplitude

Fig. 5
figure 5

Boxplots showing the percentage of the mass of a stone fragmented to < 1 mm as a function of focal pressure and transducer for burst wave lithotripsy exposures up to 30 min. The black circles are individual data points for each stone. TX1 indicates the first transducer/experiment and TX2 indicates the second transducer/experiment

In a second experiment, a separate transducer and system were used to perform BWL on 10 stones in vitro at 8 MPa pressure amplitude for up to 50 min. In this experiment, 75 ± 31% of the stone mass was reduced to < 1 mm within 30 min (Fig. 5), but only 2 stones were completely fragmented at this point (1 of which completely fragmented within 10 min). However, by 50 min’ exposure time, 9 of 10 stones were completely fragmented in this experiment, and 97 ± 0.6% of the mass was < 1 mm (Fig. 6).

Fig. 6
figure 6

Proportion of stones completely fragmented to < 1 mm pieces as a function of total exposure time in the four treatment groups across two experiments. TX1 indicates the first transducer/experiment and TX2 indicates the second transducer/experiment

In vivo evaluation in a porcine kidney and ureter

There was no evidence of gross injury in the kidneys or ureter. In one of the kidneys, there were regions of focal injury in the sinus fat and renal cortex as evidenced by bleeding (Fig. 7A). The largest region of injury was around 600 \(\mu\) m. All other focal injuries were smaller than 200 \(\mu\) m. In the ureter, there was partial sloughing of the epithelium and in regions this extended to the lamina propria (Fig. 7B). Some focal regions of damage were also observed in the muscularis (Fig. 7C). However, no hemorrhaging was observed.

Fig. 7
figure 7

A Localized bleeding in renal cortex. Total damage extends approximately 200 µm × 100 µm and is consistent with mechanical trauma due to ultrasound. B Partial dissection of the inner layer of the ureter approximately 200 µm thick. This may be from the ultrasound exposure, but the lack of any bleeding is inconsistent with previously observed injury in vivo. It may also be from stone implantation or extraction during necropsy, or histologic processing artifacts. C Focal damage and localized denudation of urothelium. Focal damage is ≤ 400 µm from the urothelial surface. Stone fragments (black) are evident in the lumen. Again, the lack of bleeding is inconsistent with previously observed mechanical injury due to ultrasound exposure. Effects are potentially from stone implantation or histologic processing artifacts

Discussion

This article described the design and testing of a BWL platform to treat ureteroliths in cats. Obstructing ureteroliths present a serious risk to renal function, as the backpressure to the pelvis can lead to severe damage if not addressed promptly [1, 7]. In small pilot trials using extracorporeal SWL in cats, improvement in azotemia was observed even in cases where no apparent stone fragmentation was detected, indicating even small changes to the stone may partially relieve obstruction [33, 34]. However, SWL is not advised for treatment in cats, as the fragmentation success is low [34, 35]. If demonstrated effective and safe, BWL could substantially improve the approach to managing obstructing ureteroliths and other stones in cats.

The studies herein demonstrate that BWL can effectively fragment feline calcium oxalate urinary stones in vitro. BWL has been primarily developed for human application and recently reported results in 19 patients [18], but there are several differences between humans and cats that necessitate adaptation of the system. First, the human ureter is approximately 3 mm intraluminal diameter and passes stones with high probability for those < 5 mm [36]. Many BWL studies have been performed at 350 kHz where nearly all fragments in calcium oxalate monohydrate stones have been found to be < 2 mm [19, 20, 37]. The feline ureter is much smaller than humans, and data suggests that fragments should be < 1 mm to maximize success [4, 11, 22]. To adjust to this difference, we exploited a unique characteristic of BWL to control the size of fragments generated, and it has been shown possible in vitro to ‘dust’ human stones to < 1 mm size with higher frequency [19]. The frequency for the present study was designed to be 700 kHz, but we found the transducer to operate at 650 kHz most efficiently. A higher frequency exposure can break smaller stones and produces smaller fragments but requires a longer time to achieve complete fragmentation in larger stones. We aim to investigate next whether a higher-frequency device would be beneficial to treating smaller stones in cats. These design methods could also be valuable in future work translating BWL to the anatomy for other cases such as pediatric human stones or canine stones.

CT analysis indicated that nearly any location along the urinary tract had an available acoustic window in the cat from the posterior flank or abdomen. The change in transducer geometry as well as focal depth minimized the volume of the focus. These changes limited the depth of field of the focus to minimize high-amplitude ultrasound exposure to surrounding organs. In practice, ultrasound imaging will also need to be performed preprocedure to ensure no sensitive organs such as intestines are present in the high-amplitude focal region. In addition, the small skin-to-stone distance required a small focus to avoid focal effects to the skin surface from cavitation. However, it is known that for both SWL and BWL, the width of the focus should be similar to or larger than the width of the stone [15, 38,39,40]. Obstructing ureteroliths in cats are commonly found to be 1–4 mm [41], thus the tradeoff was limited to treating stones in this range. Nonetheless, stones can be larger than 5 mm and these would likely be difficult to design for given the constraints and tradeoffs mentioned above.

All the stones treated in this study were composed of 100% calcium oxalate monohydrate. We did not test for the response of different stone compositions to the treatment, but a large majority of upper tract stones in cats are found to be calcium oxalate [42], and of those, calcium oxalate monohydrate is the most common type. Thus, fragmentation will possibly be more consistent than in humans, where several compositions are common. Another limitation of the model applied here was that it did not strictly represent an obstructing stone as it was positioned in a way that was open to the water bath rather than surrounded by the tissue of the ureter wall. Such mechanical differences may also impact fragmentation [43].

Previous investigations have demonstrated 350-kHz BWL can be applied with pressure amplitudes up to 7 MPa to porcine kidneys with almost no discernable injury [20]. Indeed, only minor focal damage was observed when the kidneys and ureters of pigs were treated with stones implanted, and no untreated controls were performed to assess if any of the injury was associated with the surgical implantation of the stone. It is possible that the focus was not perfectly aligned with the stone since we relied on visual and auditory cues for alignment. However, no injury was found tissues adjacent to the stone. The primary injury mechanism in lithotripsy is inertial cavitation [44, 45], which occurs from the tensile half cycles of the wave oscillating microscale bubbles in the tissue. A common metric for the potential for cavitation is the mechanical index (MI) [46], for which the MI = 11.8 for a 7 MPa, 350-kHz beam. The 650-kHz exposure at a maximum pressure of 8.9 MPa for the cat transducer results in an MI = 11.0, indicating a similar potential for cavitation to those found safe in a pig model. An additional safety measure for BWL is that cavitation is readily detected by ultrasound imaging [30], and the exposure can be temporarily terminated if detected to minimize ultrasound-related injury. With these data and additional measures, we anticipate that this transducer and BWL exposure can be safely delivered to a stone in vivo, and we plan to further evaluate this technology through future clinical testing.

Conclusions

A BWL system for noninvasive fragmentation of obstructing ureteroliths in cats was developed and tested in vitro. The transducer geometry and exposure parameters were designed to account for feline anatomic constraints common in clinical cases. Testing indicated natural calcium oxalate stones can regularly be fragmented to pieces smaller than 1 mm. These results indicate this system may be effective at fragmenting stones and relieving urinary obstruction.

Availability of data and materials

The datasets used and/or analyzed during the current study available from the corresponding author on reasonable request.

References

  1. Kyles AE, Hardie EM, Wooden BG, Adin CA, Stone EA, Gregory CR, et al. Management and outcome of cats with ureteral calculi: 153 cases (1984–2002). J Am Vet Med Assoc. 2005;226(6):937–44.

    Article  PubMed  Google Scholar 

  2. Lemieux C, Vachon C, Beauchamp G, Dunn ME. Minimal renal pelvis dilation in cats diagnosed with benign ureteral obstruction by antegrade pyelography: a retrospective study of 82 cases (2012–2018). J Feline Med Surg. 2021;23(10):892–9.

    Article  PubMed  Google Scholar 

  3. Kyles AE, Hardie EM, Wooden BG, Adin CA, Stone EA, Gregory CR, et al. Clinical, clinicopathologic, radiographic, and ultrasonographic abnormalities in cats with ureteral calculi: 163 cases (1984–2002). J Am Vet Med Assoc. 2005;226(6):932–6.

    Article  PubMed  Google Scholar 

  4. Wormser C, Clarke DL, Aronson LR. Outcomes of ureteral surgery and ureteral stenting in cats: 117 cases (2006–2014). J Am Vet Med Assoc. 2016;248(5):518–25.

    Article  PubMed  Google Scholar 

  5. Kopecny L, Palm CA, Segev G, Larsen JA, Westropp JL. Urolithiasis in cats: evaluation of trends in urolith composition and risk factors (2005–2018). J Vet Intern Med. 2021;35(3):1397–405.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Milligan M, Berent AC. Medical and interventional management of upper urinary tract uroliths. Vet Clin North Am Small Anim Pract. 2019;49(2):157–74.

    Article  PubMed  Google Scholar 

  7. Lulich JP, Berent AC, Adams LG, Westropp JL, Bartges JW, Osborne CA. ACVIM small animal consensus recommendations on the treatment and prevention of uroliths in dogs and cats. J Vet Intern Med. 2016;30(5):1564–74.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Berent AC, Weisse CW, Bagley DH, Lamb K. Use of a subcutaneous ureteral bypass device for treatment of benign ureteral obstruction in cats: 174 ureters in 134 cats (2009–2015). J Am Vet Med Assoc. 2018;253(10):1309–27.

    Article  PubMed  Google Scholar 

  9. Wuillemin F, Vachon C, Beauchamp G, Dunn M. Subcutaneous ureteral bypass device placement in 81 cats with benign ureteral obstruction (2013–2018). J Vet Intern Med. 2021;35(6):2778–86.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Kulendra NJ, Borgeat K, Syme H, Dirrig H, Halfacree Z. Survival and complications in cats treated with subcutaneous ureteral bypass. J Small Anim Pract. 2021;62(1):4–11.

    Article  CAS  PubMed  Google Scholar 

  11. Deroy C, Rossetti D, Ragetly G, Hernandez J, Poncet C. Comparison between double-pigtail ureteral stents and ureteral bypass devices for treatment of ureterolithiasis in cats. J Am Vet Med Assoc. 2017;251(4):429–37.

    Article  PubMed  Google Scholar 

  12. Rassweiler JJ, Knoll T, Kohrmann KU, McAteer JA, Lingeman JE, Cleveland RO, et al. Shock wave technology and application: an update. Eur Urol. 2011;59(5):784–96.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Rassweiler J, Rassweiler MC, Klein J. New technology in ureteroscopy and percutaneous nephrolithotomy. Curr Opin Urol. 2016;26(1):95–106.

    Article  PubMed  Google Scholar 

  14. Cléroux A. Minimally invasive management of uroliths in cats and dogs. Vet Clin North Am Small Anim Pract. 2018;48(5):875–89.

    Article  PubMed  Google Scholar 

  15. Maxwell AD, MacConaghy B, Bailey MR, Sapozhnikov OA. An investigation of elastic waves producing stone fracture in burst wave lithotripsy. J Acoust Soc Am. 2020;147(3):1607.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Sorensen MD, Harper JD, Hsi RS, Shah AR, Dighe MK, Carter SJ, et al. B-mode ultrasound versus color Doppler twinkling artifact in detecting kidney stones. J Endourol. 2013;27(2):149–53.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Cunitz BW, Harper JD, Sorensen MD, Haider YA, Thiel J, May PC, et al. Quantification of renal stone contrast with ultrasound in human subjects. J Endourol. 2017;31(11):1123–30.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Harper JD, Lingeman JE, Sweet RM, Metzler IS, Sunaryo PL, Williams JC, et al. Fragmentation of stones by burst wave lithotripsy in the first 19 humans. J Urol. 2022;207(5):1067–76.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Maxwell AD, Cunitz BW, Kreider W, Sapozhnikov OA, Hsi RS, Harper JD, et al. Fragmentation of urinary calculi in vitro by burst wave lithotripsy. J Urol. 2015;193(1):338–44.

    Article  PubMed  Google Scholar 

  20. Maxwell AD, Wang Y-N, Kreider W, Cunitz BW, Starr F, Lee D, et al. Evaluation of renal stone comminution and injury by burst wave lithotripsy in a pig model. J Endourol. 2019;33(10):787–92.

    Article  PubMed  PubMed Central  Google Scholar 

  21. May PC, Kreider W, Maxwell AD, Wang YN, Cunitz BW, Blomgren PM, et al. Detection and evaluation of renal injury in burst wave lithotripsy using ultrasound and magnetic resonance imaging. J Endourol. 2017;31(8):786–92.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Berent AC, Weisse CW, Todd K, Bagley DH. Technical and clinical outcomes of ureteral stenting in cats with benign ureteral obstruction: 69 cases (2006–2010). J Am Vet Med Assoc. 2014;244(5):559–76.

    Article  PubMed  Google Scholar 

  23. Legget ME, Leotta DF, Bolson EL, McDonald JA, Martin RW, Li XN, et al. System for quantitative three-dimensional echocardiography of the left ventricle based on a magnetic-field position and orientation sensing system. IEEE Trans Biomed Eng. 1998;45(4):494–504.

    Article  CAS  PubMed  Google Scholar 

  24. Kim Y, Maxwell AD, Hall TL, Xu Z, Lin KW, Cain CA. Rapid prototyping fabrication of focused ultrasound transducers. IEEE Trans Ultrason Ferroelectr Freq Control. 2014;61(9):1559–74.

    Article  PubMed  Google Scholar 

  25. Blackstock DT. Fundamentals of Physical Acoustics. New York, NY: Wiley; 2000.

    Google Scholar 

  26. Fan X, Hynynen K. The effect of wave reflection and refraction at soft tissue interfaces during ultrasound hyperthermia treatments. J Acoust Soc Am. 1992;91(3):1727–36.

    Article  CAS  PubMed  Google Scholar 

  27. Maréchal P, Levassort F, Tran-Huu-Hue L-P, Lethiecq M. Lens-focused transducer modeling using an extended KLM model. Ultrasonics. 2007;46(2):155–67.

    Article  PubMed  Google Scholar 

  28. Maxwell AD, Yuldashev PV, Kreider W, Khokhlova TD, Schade GR, Hall TL, et al. A prototype therapy system for transcutaneous application of boiling histotripsy. IEEE Trans Ultrason Ferroelectr Freq Control. 2017;64(10):1542–57.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Wurster C, Köhler M, Pecha R, Eisenmenger W, Suhr D, Irmer U, et al. Negative pressure measurements of water using the glass fiber optic hydrophone. Proc 1st World Congr on Ultrasonics (Berlin). 1995;1(1):635–8.

  30. Maxwell AD, Hunter C, Cunitz BW, Kreider W, Totten S, Wang YN. Factors affecting tissue cavitation during burst wave lithotripsy. Ultrasound Med Biol. 2021;47(8):2286–95.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Harper JD, Metzler I, Hall MK, Chen TT, Maxwell AD, Cunitz BW, et al. First in-human burst wave lithotripsy for kidney stone comminution: initial two case studies. J Endourol. 2021;35(4):506–11.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Hall MK, Thiel J, Dunmire B, et al. Feasibility study of using point of care ultrasonic propulsion and burst wave lithotripsy (BWL) to noninvasively treat symptomatic ureteral stones. J Urol. 2022;208:1075–82.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Defarges A, Berent A, Dunn M. New alternatives for minimally invasive management of uroliths: ureteroliths. Compend Contin Educ Vet. 2013;35(3):E4.

    PubMed  Google Scholar 

  34. Adams LG, Senior DF. Electrohydraulic and extracorporeal shock-wave lithotripsy. Vet Clin North Am Small Anim Pract. 1999;29(1):293–302.

    Article  CAS  PubMed  Google Scholar 

  35. Adams LG, Williams JC, McAteer JA, Hatt EK, Lingeman JE, Osborne CA. In vitro evaluation of canine and feline calcium oxalate urolith fragility via shock wave lithotripsy. Am J Vet Res. 2005;66(9):1651–4.

    Article  PubMed  Google Scholar 

  36. Coll DM, Varanelli MJ, Smith RC. Relationship of spontaneous passage of ureteral calculi to stone size and location as revealed by unenhanced helical CT. AJR Am J Roentgenol. 2002;178(1):101–3.

    Article  PubMed  Google Scholar 

  37. Zwaschka TA, Ahn JS, Cunitz BW, Bailey MR, Dunmire B, Sorensen MD, et al. Combined burst wave lithotripsy and ultrasonic propulsion for improved urinary stone fragmentation. J Endourol. 2018;32(4):344–9.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Randad A, Ghanem MA, Bailey MR, Maxwell AD. Design, fabrication, and characterization of broad beam transducers for fragmenting large renal calculi with burst wave lithotripsy. J Acoust Soc Am. 2020;148(1):44.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Neisius A, Smith NB, Sankin G, Kuntz NJ, Madden JF, Fovargue DE, et al. Improving the lens design and performance of a contemporary electromagnetic shock wave lithotripter. Proc Natl Acad Sci U S A. 2014;111(13):E1167–75.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  40. Sapozhnikov OA, Maxwell AD, MacConaghy B, Bailey MR. A mechanistic analysis of stone fracture in lithotripsy. J Acoust Soc America. 2007;121(2):1190–202.

    Article  Google Scholar 

  41. Nesser VE, Reetz JA, Clarke DL, Aronson LR. Radiographic distribution of ureteral stones in 78 cats. Vet Surg. 2018;47(7):895–901.

    Article  PubMed  Google Scholar 

  42. Cannon AB, Westropp JL, Ruby AL, Kass PH. Evaluation of trends in urolith composition in cats: 5,230 cases (1985–2004). J Am Vet Med Assoc. 2007;231(4):570–6.

    Article  CAS  PubMed  Google Scholar 

  43. Randad A, Ahn J, Bailey MR, Kreider W, Harper JD, Sorensen MD, et al. The impact of dust and confinement on fragmentation of kidney stones by shockwave lithotripsy in tissue phantoms. J Endourol. 2019;33(5):400–6.

    Article  PubMed  PubMed Central  Google Scholar 

  44. Zhong P, Zhou Y, Zhu S. Dynamics of bubble oscillation in constrained media and mechanisms of vessel rupture in swl. Ultrasound Med Biol. 2001;27(1):119–34.

    Article  CAS  PubMed  Google Scholar 

  45. Evan AP, Willis LR, McAteer JA, Bailey MR, Connors BA, Shao Y, et al. Kidney damage and renal functional changes are minimized by waveform control that suppresses cavitation in shock wave lithotripsy. J Urol. 2002;168(4 Pt 1):1556–62.

    Article  PubMed  Google Scholar 

  46. Apfel RE, Holland CK. Gauging the likelihood of cavitation from short-pulse, low-duty cycle diagnostic ultrasound. Ultrasound Med Biol. 1991;17(2):179–85.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was financially supported by National Institutes of Health NIDDK P01 DK043881, NIDDK K01 DK104854, ORIP K01 OD019912, and the EveryCat Health Foundation. The funding bodies played no role in the design of the study and collection, analysis, interpretation of data, and in writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

A.D.M., G.W.K., B.M., E.L., and M.R.B. built and tested the device. E.F., J.L., M.T., M.S.B., and Y.N.W., provided design specifications and tested safety. D.F.L. led the processing of computed tomography images. All authors contributed to writing the manuscript and preparing figures.

Corresponding author

Correspondence to Michael R. Bailey.

Ethics declarations

Ethics approval and consent to participate

We confirm that all experimental protocols were approved by the University of Washington IACUC.

We confirm that all methods were carried out in accordance with relevant guidelines and regulations.

We confirm that all methods are reported in accordance with ARRIVE guidelines ([arriveguidelines.org] https://arriveguidelines.org [arriveguidelines.org]) for the reporting of animal experiments.

Euthasol (euthanasia solution), 390 mg/ml, 100 ml bottle is the chemical used in the euthanasia process during the study.

The animals were obtained with the consent of the vendor to use them for research.

Consent for publication

This section is not applicable. The manuscript does not include details, images, or videos relating to an individual person.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Maxwell, A.D., Kim, G.W., Furrow, E. et al. Development of a burst wave lithotripsy system for noninvasive fragmentation of ureteroliths in pet cats. BMC Vet Res 19, 141 (2023). https://doi.org/10.1186/s12917-023-03705-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12917-023-03705-1