Standardization of implantation technique
All implantations were carried out by the same experimenter, who was skilled and experienced in microsurgery and implantation of devices in mice for long-term survival experiments. In a preliminary pilot experiment (see below), the implantation methodology (including supportive measures during implantation) was established and practiced until the technique was optimized and highly reproducible.
Implantation of blood pressure transmitters
The implantations were carried out in a laminar flow hood equipped with a surgical microscope. Aseptic conditions were assured by the use of autoclaved instruments and sterilized materials. Inhalation anaesthesia was initiated and maintained, respectively, with 7–8% and 3.5–4% sevoflurane (Sevorane®, Abbot, Cham, Switzerland) in O2. During anaesthesia, the animal's eyes were protected with ointment (Vitamine A, Bausch & Lomb, Steinhausen, Switzerland). After shearing of hairs and disinfection of the anterior neck region, a longitudinal skin incision of 1 cm was performed, the connective tissues and muscles were prepared and the left common carotid artery (Arteria carotis communis sinistra) was exposed. The artery was ligated with a silk suture (PERMA-Handseide, 6-0, Ethicon, Norderstedt, Germany) at its bifurcation at both the internal and external branches. A second silk suture was placed around the artery at a distance of 4–6 mm caudal of the bifurcation and the blood flow was stopped by retracting the suture. A third silk suture was bound loosely around the artery between the other two. Then, a hole was cut in the artery using fine-bladed scissors and the catheter of a TA11PA-C10 transmitter (DataSciences International, St. Paul, MN, USA) was inserted into the vessel. The middle silk suture was tightened and the retracting, caudal thread was released to allow the tip of the catheter to be advanced into the thoracic aorta. The catheter was then secured in the carotid artery by knotting the silk sutures. The muscles and connective tissues were restored with absorbable sutures (VICRYL 6-0, Ethicon, Norderstedt, Germany). A pocket between the skin and the muscle layers was prepared by blunt dissection with atraumatic scissors. This pocket in the subcutaneous connective tissues reached from the right edge of the skin incision to the back of the mouse. The transmitter body was inserted into the pocket at the right edge of the skin incision and then advanced in a dorsal direction until it was located subcutaneously on the upper back. The transmitter body was then fixed with 3–4 loops of surgical stainless steel sutures (EH 7623G, 3-0; Ethicon, Norderstedt, Germany), which were laid from one side to the other through the skin in the subcutaneous connective tissues underneath the transmitter body. One to two loops of stainless steel wire were placed behind the transmitter body to prevent its dislocation to the lower back. Finally, the skin incision in the neck was closed with absorbable sutures and anaesthesia was stopped. Throughout the entire study, the time required for anaesthesia and implanting the transmitter in each individual was routinely 50–60 min.
Supportive measures during implantation
The laminar flow work bench on which the mouse was laid during surgery was equipped with a water-bath-heated surface set at 38°C to avoid any drop in the mouse's body temperature due to anaesthesia. When anaesthesia was withdrawn, the conscious animal was allowed to recover for 2–3 hours on the warmed surface (during which time it could move freely under a wire mesh grid) before being transferred back to its home cage.
To support fluid homeostasis during surgery, 1 mL of saline (0.9%) at a temperature of 36°C was injected intraperitoneally shortly after induction of anaesthesia. This was also intended to prevent hypovolemia and hypotension in the case of substantial blood loss from possible bleeding during surgery.
As anti-infective prophylaxis, Sulfadoxin and Trimethoprim were injected intraperitoneally at a dosage of 30 and 6 mg/kg bodyweight, respectively, before starting surgery.
Analgesics
Either buprenorphine (Temgesic®, 0.1 mg/kg body weight; Essex Chemie AG, Lucerne, Switzerland) or flunixine (Biokema Flunixine®, 5 mg/kg body weight; Biokema SA, Crissier-Lausanne, Switzerland) was applied. Analgesics were administered subcutaneously twice per day (i.e. every 12 hours) for 7 days; the first injection of analgesics was performed during transmitter implantation.
Postoperative analgesia and care regimens
All animals recovered within 10–15 min after cessation of inhalation anaesthesia, and showed undisturbed moving behaviour, self grooming, and occasional climbing at 1–2 hours after implantation. All mice were active and in good general condition when transferred to their home cage 2–3 hours after implantation was completed.
Mice in the first two groups were transferred back to standard housing conditions (described below) 2–3 hours after implantation was complete. In the first group, pain was alleviated with buprenorphine (wt: n = 2, tg6: n = 2); in the second group flunixine was used as analgesic (wt: n = 1, tg6: n = 4). No further postoperative measures were applied in these animals. All animals from these two groups were found moribund with severe symptoms of hypothermia, apathy, and exsiccosis at 1–14 days after implantation. The wt mice in the first and second group had a mean survival time of 9 days, whereas the tg6 mice had a mean survival time of 4 days after implantation. Animals treated with the opioid buprenorphine (first group) had a mean survival time of 4 days, whereas treatment with the non-steroidal anti-inflammatory drug flunixine (second group) showed a tendency towards longer survival, with a mean survival time of 7 days.
Necropsy and transmitter explantation both confirmed that the implantation per se was optimal, with no signs of specific complications (e.g., bleeding, necrosis, inflammation, wound dehiscence, or entrapment of legs or teeth in a wire loop). Further systematic pathological investigation revealed no abnormalities and no hints of side effects from the different types of pain killers used. It was assumed that the prolonged deficiency in energy and fluid intake combined with the extended postoperative energy demand and temperature loss were the causes underlying the deterioration in bodily functions and fatal outcome in the first two groups.
The results from the first two groups led us to develop improvements in the postoperative care regimen focussed on resolving the energy and fluid deficiencies and supporting maintenance of normal body temperature.
A third group (wt: n = 47, tg6: n = 48) received flunixine for pain relief and an additional fluid therapy of 300 μl glucose (5%) and 300 μl saline (0.9%), injected subcutaneously twice per day for 7 days. All analgesic agents including glucose liquid and saline were warmed to body temperature before injection. For this third group, all animal cages were kept on a heating mat during the entire 2-week postoperative period. In addition, unlike the first two groups, the third group had free access to glucose (15%), offered in a second drinking bottle, and to a high-energy wet food (Solid Drink®-Energy, Triple A Trading, Tiel, Netherlands). The high-energy food and the glucose-containing bottle were provided in the animals' cages from 2 to 3 days before surgery until 2 weeks afterwards. In addition, O2 was piped into the cages for two weeks after implantation.
In contrast to the animals of the first and second group, no animal in the third group showed postoperative complications such as hypothermia, apathy, or exsiccosis. Four tg6 mice developed haematoma in the right neck and shoulder area at 1–2 hours after surgery. Three of these animals could be rescued by re-opening the wound and removing the blood clots, while additional warm saline (1 mL) was injected intraperitoneally. Despite removing the haematoma, one of these animals was sacrificed 6 days after implantation because of repeated bleeding at the implantation site. Finally, 94 out of 95 mice in group 3 were healthy and in good physical condition at the end of the 14-day postoperative period.
Later, three wt mice were sacrificed between 23 and 30 days after implantation as these animals were fighting with their female companions and showed severe skin injuries. Two tg6 mice were sacrificed 23 and 41 days after implantation because the telemetric signals hinted at signs of blood clot formation at the tip of the catheter (i.e. damped blood pressure curves). This was confirmed during necropsy. Despite these limitations, the methodology applied to group 3 was rather successful, as shown by the relationship between postoperative treatment and survival time (Figure 1). The postoperative intensive care regimen established in the third group was associated with survival times of 30–56 days, which was the final endpoint of the study and the point at which the animals were sacrificed after having exercised on the treadmill.
Survival time vs. body weight
Animals underwent implantation of telemetric transmitters at the age of 4–5 weeks (mean 30.9 days, standard deviation ± 2.7 days). Mice were weighed on the day of implantation (before implantation started) and on the day on which they were sacrificed. Body weight was recorded with a precision balance (PR 2003 Delta Range, Mettler-Toledo AG, Greifensee, Switzerland) especially calibrated for use with moving animals. Body weights were corrected to take into account the weight of the transmitter (1.4 g). All weighing procedures were performed at the same time of day (07:00–08:00 a.m.).
On the day of implantation, body weights ranged from 18 g to 27 g (wt: mean 21.6, standard deviation ± 1.6; tg6: mean 22.2, standard deviation: ± 2.1). There was no correlation between an individuals' body weight at implantation and their survival time. Mice that did not survive more than 2 weeks were around average body weight, or even slightly above-average weight, at implantation (Figure 2). The final body weight of mice that did not survive more than 2 weeks was decreased in the tg6 mice but not in wt animals. Thus, the final body weight was not clearly correlated with survival time. The final body weight of the 94 mice that survived >3 weeks showed that they gained weight over time, as would be expected with increasing age and bodily development (Figure 3).
Long-term effects of transmitter fixation
In all animals, the transmitter body was fixed in the midline of the mouse's back (Figure 4). In two animals from the third group, the transmitter body had to be re-fixed the day after implantation because it had turned and moved to the upper neck region, where it could hamper the movement of the mouse's head. In general, the mice showed no signs of reduced physical activity or restricted head movement. In contrast to the implantation technique usually used, where the transmitter body hangs loosely at the lateral body side, fixation with stainless steel wires on the animal's back allowed reliable detection of the telemetric signal, as the distance between the telemetric receiver plate and the transmitter body is shortened (Figure 5).
Telemetric signal verification during maximal exercise
The reliability of constant telemetric signaling during maximal exercise was demonstrated in 89 mice at 30–56 days after implantation (Figure 6). The telemetric signal was measured during maximal exercise using a Simplex II metabolic rodent treadmill, fitted with an Oxymax gas analyzer (Columbus Instruments, Columbus, OH, USA). For the maximal incremental exercise test, mice were placed in the exercise chamber and allowed to equilibrate for 30 min. Treadmill activity was initiated at 2.5 m/min and 0° inclination for 10 min, and then increased by 2.5 m/min and 2.5° every 3 min thereafter until exhaustion. Mice were gently encouraged to run for as long as possible by the use of a mild electric grid at the end of the treadmill (0.2 mA, pulse 200 ms, 1 Hz). Exhaustion was defined as the inability to continue regular treadmill running despite a repeated electric stimulus to the mice. Accurate arterial blood pressure curves with minimal artefacts were recorded during maximal exercise on the treadmill in all mice tested.