Skip to main content

Table 2 Measured SAA and observed/expected ratios following addition of interfering substances to feline samples with low-moderate and moderate-high SAA concentrations

From: Validation of an automated immunoturbidimetric assay for feline serum amyloid A

Interferent

Interferent Concentration

Measured SAA (mg/L)

Low-Moderate

O/E (%)b

Measured SAA (mg/L)

Moderate-High

O/E (%)b

Haemoglobin

Blank

18.37

100

49.83

100

0.146 g/L

18.67

103.02

51.33

103.02

1.46 g/L

18.28

101.28

50.47

101.28

14.6 g/L

17.2

96.35

48.01

96.35

Triglyceride

Blank

18.37

100

49.83

100

2.74 mmol/L

17.95

97.7

53.71

107.81

4.91 mmol/L

19.44

105.81

51.27

102.91

9.57 mmol/L

17.4

94.7

51.41

103.19

Bilirubin

Blank

18.37

100

49.83

100

27 μmol/L

24.57

133.74a

52.42

105.2

46 μmol/L

22.4

121.92a

52.69

105.75

96 μmol/L

19.61

106.77

50.11

100.57

  1. aIndicates bias above acceptable limit of ±10% of the blank measurement. b Observed/Expected (%)