Table 2 Trending ability with concordance and polar plot data for 18 anaesthetized rabbits

From: Prospective clinical study to evaluate an oscillometric blood pressure monitor in pet rabbits

  Systolic Mean Diastolic
Concordance rate
 Exclusion zone 5% 67% 84% 67%
 Exclusion zone 10% 65% 87% 72%
 Polar plot analysis (n) 82 100 90
Mean polar angle
 Exclusion zone 5% −13.7° −9.5° − 12.0°
 Standard deviation 28.0° 28.1° 31.4°
Radial limits of agreement
 Exclusion zone 5% − 56.4° to 42.2° −65.1° to 46.6° −70.2° to 50.8°
  1. Data points representing percentage changes lower than 5% (exclusion zone 5%) or 10% (exclusion zone 10%) were not included in the analysis. n: number of observations
\