Table 1 Mean, standard deviation (SD), intra- and inter-assay CVs variation in TACc concentrations of canine serum samples (two with lower concentration and two with high concentrations) measured in the two apparatus

From: Validation of an automated assay for the measurement of cupric reducing antioxidant capacity in serum of dogs

Apparatus Validation parameter Number of samples Mean (mmol Trolox equiv./L) SD CV (%)
Cobas Mira Plus Intra-assay 2 0.2325 0.004 1.9
2 0.3823 0.007 1.8
Inter-assay 2 0.2254 0.015 6.8
2 0.3766 0.019 5.1
Olympus AU400 Intra-assay 2 0.2623 0.003 1.2
2 0.3999 0.003 0.7
Inter-assay 2 0.2533 0.022 8.6
2 0.3842 0.030 7.8
  1. TAC c total antioxidant capacity using the cupric reducing antioxidant capacity methodology and the bathocuproinedisulfonic acid disodium salt, SD standard deviation, CV coefficient of variation