Skip to main content

Table 1 Mean, standard deviation (SD), intra- and inter-assay CVs variation in TACc concentrations of canine serum samples (two with lower concentration and two with high concentrations) measured in the two apparatus

From: Validation of an automated assay for the measurement of cupric reducing antioxidant capacity in serum of dogs

Apparatus

Validation parameter

Number of samples

Mean (mmol Trolox equiv./L)

SD

CV (%)

Cobas Mira Plus

Intra-assay

2

0.2325

0.004

1.9

2

0.3823

0.007

1.8

Inter-assay

2

0.2254

0.015

6.8

2

0.3766

0.019

5.1

Olympus AU400

Intra-assay

2

0.2623

0.003

1.2

2

0.3999

0.003

0.7

Inter-assay

2

0.2533

0.022

8.6

2

0.3842

0.030

7.8

  1. TAC c total antioxidant capacity using the cupric reducing antioxidant capacity methodology and the bathocuproinedisulfonic acid disodium salt, SD standard deviation, CV coefficient of variation