Skip to main content

Table 1 Imprecision study for the STAGO STA-Liatest D-di + (Stago) and NycoCard.™ D-dimer (NycoCard) assays. Intra- and inter-coefficients of variation (CV) for the low and medium pool for the Stago and NycoCard were calculated as CV % = (standard deviation / mean) * 100

From: Investigation of two different human d-dimer assays in the horse

 

Stago

NycoCard

Intra-CV % Low pool

4.66% (0.23; 0.01)

35.14% (0.15; 0.053)

Inter-CV % Low pool

5.98% (0.21; 0.01)

33.99% (0.13; 0.04)

Intra-CV % Medium pool

1.79% (0.51; 0.01)

11.11% (0.60; 0.07)

Inter-CV % Medium pool

2.63% (0.49; 0.01)

12.33% (0.58; 0.07)

  1. (mean; standard deviation)