Skip to main content

Table 2 Analytical inaccuracy of serum amyloid A (SAA) concentrations assessed by dilution of an equine serum pool

From: Validation of an equine serum amyloid A assay with an unusually broad working range

Full concentrations range (up to 6159.7 mg/L)y-intercept (95% confidence interval)− 130 (−293.3, 33.19)
Slope (95% confidence interval)1.087 (1.035, 1.140)
P (Runs test)0.0002
r20.99
Concentration range up to 3000 mg/Ly-intercept (95% confidence interval)− 58.5(− 129.6, 12.59)
Slope (95% confidence interval)0.952 (0.902, 1.003)
P (Runs test)0.4
r20.99