Skip to main content

Table 1 Intra- and inter-assay variation in determination of serum amyloid (SAA) concentrations in equine serum samples

From: Validation of an equine serum amyloid A assay with an unusually broad working range

ComparisonNo. of repeatsSAA concentration (mg/L)Coefficient of variation (%)
MeanSD
Intra-assay
 Low conc.10124.73.83.0
 Intermediate conc.10805.341.95.2
 High conc.104443161.33.6
Inter-assay
 Low conc.10113.47.76.8
 Intermediate conc.10842.063.67.6
 High conc.104646445.79.6