Table 7 Secondary dose verification to the surface of the skin at the time of the first delivery, nanoDot™ dosimeters were positioned in areas where classical TSEBT uncertainties would present or at mold intersections where a gap may exist

From: A case report of total skin photon radiation therapy for cutaneous epitheliotropic lymphoma in a dog

Site

NanoDot Serial Number

Surface Dose (cGy)/Fraction

%D Prescription Dose

Dorsal Head (Vertex)

DN086200581

180.684

0.38

Dorsal Thorax (T6-T7)

DN08620795P

180.991

0.55

Dorsal Abdomen (Sacrum)

DB08305757G

180.584

0.32

Ventral Neck (Larynx)

DN09050308Y

176.349

−2.03

Ventral Thorax (Xiphoid)

DN08749836D

179.259

−0.41

Ventral Abdomen (Inguinal)

DN08749729A

180.181

0.10

  1. Prescription Dose: 180.00 cGy. Uncertainty of NanoDot measurement: +/− 1.7%