Skip to main content

Table 4 MR sequences and imaging parameters used during the study

From: Magnetic resonance imaging of liver tumors using gadoxetic acid (Gd-EOB-DTPA) - pilot study

 

Pre-contrast

T1-W TFE

T2-W MV HR

Post-contrast

T1-W TFE

Echo time (ms)

4.6

100

4.6

Repetition time (ms)

10

3745

10

Flip angle (°)

25

90

25

Bandwidth (Hz)

141.7

531.9

141.7

FOV (mm)

220 × 192

250 × 250

220 × 192

Matrix

220 × 192

500 × 500

220 × 192

Slice thickness (mm)

3

5

3

GAP (mm)

0

0.5

0

Number of slices

100

50

100