Table 2 Implant acceptance criteria

From: A retrospective review of 146 active and passive fixation bradycardia lead implantations in 74 dogs undergoing pacemaker implantation in a research setting of short term duration

Parameters Atrium Right Ventricle
Threshold @ -0.5 ms PW ≤ 1.5 V ≤ 1.0 V
Sensing (amplitude) ≥ 2.0 mV ≥ 5.0 mV
Impedance (2.5 V or 5 V, 0.6 ms) 300–1800 Ω 300–1800 Ω