From: Development of a novel lateral flow assay for detection of African swine fever in blood

Fig. 2

Comparison of the performance of the LFA and the commercial DAS-ELISA. Serial dilutions of semi-purified VP72 protein in blood were tested by LFA and Double antibody sandwich ELISA (DAS-ELISA) and results were compared. Both assays are based on MAbs specific for VP72 protein. The detection limit of both tests was very similar (3 ng viral protein/test by LFA vs 1,9 ng viral protein/test in ELISA)