Table 2 Correlation and differences between the different gates for canine r-PLT measurement (n = 362) of the Sysmex XT-2000iV

From: Optimized gating and reference ranges of reticulated platelets in dogs for the Sysmex XT-2000iV

Variable rs Slope (95 % CI) Intercept (95 % CI) Bias (95 % limits of agreement) Adjusted P-value
r-PLT “Gelain-gate”(%) vs. r-PLT “Oellers-gate”(%) 1.00 1.01 (1.01–1.01) 0.00 (0.00–0.00) 0.1 (−1.3–1.6) ns
r-PLT “Pankraz-gate”(%) vs. r-PLT “Gelain-gate”(%) 0.89 2.83 (2.70–2.99) 0.18 (0.12–0.23) 2.0 (−4.3–8.3) <0.0001
r-PLT “Pankraz-gate”(%) vs. r-PLT “Oellers-gate”(%) 0.88 2.92 (2.78–3.11) 0.17 (0.10–0.22) 2.1 (−4.9–9.2) <0.0001
  1. Abbreviations: CI confidence interval, ns not significant, vs. versus