Table 2 Precision of the hscTnT assay (Roche Diagnostics, Cobas-e601) for measuring equine plasma

From: Analytical validation and establishment of reference intervals for a ‘high-sensitivity’ cardiac troponin-T assay in horses

  Level 1 (L1) Level 2 (L2) Level 3 (L3)
Within-Run hscTnT
Mean (CV), ng/L
6.5 (14.9 %) 10.1 (8.7 %) 15.3 (5.4 %)
Between-Run hscTnT Mean (CV), ng/L 12.2 (8.4 %) 57.0 (8.4 %) 256.0 (9.0 %)
  1. Abbreviations: CV Coefficient of variation (in %)
  2. Precision of the hscTnT assay (Roche Diagnostics, Cobas-e601) for measuring equine plasma hscTnT of three increasing levels (Level 1, 2, 3) on 10 separate occasions