Skip to main content

Table 3 Thresholds and corresponding test characteristics of 3 different hand-held meters to detect hyperglycemia in capillary and coccygeal blood based on Receiver Operating characteristics (ROC) analysis using a laboratory plasma glucose concentration of ≥ 60 mg/dL as reference

From: Glucose concentration in capillary blood of dairy cows obtained by a minimally invasive lancet technique and determined with three different hand-held devices

Devicea

Capillary blood

Coccygeal blood

 

Optimized device thresholdb [mg/dL]

AUCc [%(CI95 d)]

See [%(CI95)]

Spf [%(CI95)]

Youden-Indexg

Optimized device thresholdc [mg/dL]

AUC [%(CI95)]

Se [%(CI95)]

Sp [%(CI95)]

Youden-Index

FSP

43

87.4 (82.5–91.3)

76.4 (66.2–84.8)

84.1 (77.3–89.5)

0.61

59

91.7 (87.5–94.9)

75.3 (65–83.8)

91.4 (85.7–95.3)

0.67

GLX

49

93.4 (89.5–96.2)

92.1 (84.5–96.8)

85.4 (78.8–90.6)

0.78

49

85.2 (80.1–89.5)

85.4 (76.3–92)

69.5 (61.5–76.8)

0.55

WGC

95

70.5 (64.3–76.3)

39.3 (29.1–50.3)

92.1 (86.5–95.8)

0.31

94

70.0 (63.8–75.7)

55.1 (44.1–65.6)

84.8 (78–90.1)

0.40

  1. aFSP: FreeStyle Precision, (Abbott GmbH & Co. KG, Wiesbaden, Germany), GLX: GlucoMen LX Plus (A. Menarini GmbH, Vienna, Austria), WGC: WellionVet GLUCO CALEA (MED TRUST Handelsges.m.b.H, Marz, Austria)
  2. bbased on Receiver Operating Characteristics (ROC) analyses
  3. cAUC: area under the receiver operating characteristics (ROC) curve
  4. dCI95: 95 % confidence interval
  5. eSe: sensitivity
  6. fSp; specificity
  7. gYouden-Index = (Se + Sp – 1)