Figure 2 | BMC Veterinary Research

Figure 2

From: Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate

Figure 2

Testing results between long and short plate working length stabilization techniques. Comparison of mean ± SD values of construct axial stiffness (N/mm) loaded to body weight between the long plate working length stabilization technique (□) and the short plate working length stabilization technique (■). Measurements recorded at 1000, 2000, 5000, 10000, 20000, 50000, 100000, 120000, and 180000 cycles. No significant difference was found between the two stabilization techniques at any of the evaluated cycles.