Skip to main content

Table 6 Comparison of precision for TNCC between automated Sysmex BF-TNCC after regating and manual counting with hemocytometer

From: Comparison of Sysmex XN-V body fluid mode and deep-learning-based quantification with manual techniques for total nucleated cell count and differential count for equine bronchoalveolar lavage samples

Method

Sample cellularity

Mean TNCC (cells × 106/µL)

n

SD (cells × 106/µL)

CV (%)

Sysmex BF

Low

76.8

10

3.39

4.4

Moderate

263.4

10

26.34

11.8

High

714.2

10

22.31

3.1

Manual

Low

42.5

10

26.48

62.3

Moderate

252.5

10

49.23

19.4

High

997.5

10

205.29

20.6

  1. BF Body fluid, CV Coefficient of variation, n Number of replicates, SD Standard deviation, TNCC Total nucleated cell count